RE: [fluka-discuss]: USRBIN DOSE-EQ units

From: Vasilis Vlachoudis <Vasilis.Vlachoudis_at_cern.ch>
Date: Mon, 25 Apr 2016 06:44:35 +0000

Dear Nkanyiso,

all quantities in the decay part of the calculation are expressed per unit time pSv/s.
In the prompt part of the calculation are expressed as pSv/primary

Best
Vasilis

________________________________
From: owner-fluka-discuss_at_mi.infn.it [owner-fluka-discuss_at_mi.infn.it] on behalf of NKANYISO ROBERT Mantengu [3477531_at_myuwc.ac.za]
Sent: Sunday, April 24, 2016 16:03
To: fluka-discuss_at_fluka.org
Cc: Joachim Vollaire
Subject: [fluka-discuss]: USRBIN DOSE-EQ units

Dear FLUKA users,

I have a confusion with the units of the USRBIN DOSE-EQ scoring detector.

I have set the USRBIN DOSE-EQ scoring detector associated with DCYSCORE in Semi-analogue mode since I want to score prompt-decay particles, and I have also defined the irradiation profile using IRRPROFIle card.

So I'm confused if my units for DOSE-EQ should be in pSv per unit primary weight, since my DCYSCORE is in Semi-analogue mode, or they are expressed in pSv/s, since I have define the irradiation profile.

And if the units are expressed in pSv per unit primary weight, can I just multiply with my beam current (in prim/s) to get pSv/s, or I should multiply with the total particle statistical weight, which is the number of events requested, to get pSv.

Any suggestion will be appreciated.
Thanks in advance

Regards,
Nkanyiso R. Mantengu.


__________________________________________________________________________
You can manage unsubscription from this mailing list at https://www.fluka.org/fluka.php?id=acc_info
Received on Mon Apr 25 2016 - 10:47:58 CEST