Re: [fluka-discuss]: FLUKA run Block with Gamma

From: <alfredo.ferrari_at_mi.infn.it>
Date: Thu, 3 Jun 2021 21:40:56 +0200

Dear Alessandro

you put in what(4) of the detect card 2D-4, while I suspect you intended
it as the maximum value of the histogram -> it should have been in
what(3).

Since there is no default for what(3), you were inadvertently requesting a
"negative" histogram with "minimum" 1D-10 and maximum 0, with a consequent
array out of bound error triggered by that situation.
 Move the 2D-4 into what(3) and everything will work!

Ciao
Alfredo


> Dear FLUKA experts,
>
> I am doing some test with a silicon detector to study its response as a
> function of the photon energy imping on it. Unfortunately I found out a
> problem. The run start with no problem but at some point it stop and the
> run freeze. Flair says that 2 seconds remains, for example, but the
> timer remains blocked and the run did not go further.
>
> I attach the input on the email. It is a very simple gemetry with only
> the Si detector. I put the EM-Cut card but I think tha I am making some
> mistakes with the physics cards requested to transport the photons
> correctly.
>
> Which cards do I have to add? Do I have to choose a different Default
> option?
>
> Best regards and thanks for your time, Alessandro Calamida.
>
>
>
> --
> Questa e-mail รจ stata controllata per individuare virus con Avast
> antivirus.
> https://www.avast.com/antivirus
>


__________________________________________________________________________
You can manage unsubscription from this mailing list at https://www.fluka.org/fluka.php?id=acc_info
Received on Thu Jun 03 2021 - 23:40:31 CEST

This archive was generated by hypermail 2.3.0 : Thu Jun 03 2021 - 23:40:33 CEST