Re: Ionisation losses fluctuations

From: <Andrea.Mairani_at_mi.infn.it>
Date: Sat, 21 Aug 2010 10:01:03 +0200 (CEST)

Dear Antonello,
can you please send us your input?
Looking at your USRBIN,
  I see that you chose as the output unit 19 (use units> 21), in fact the
manual clearly states:
   " WHAT(3) = logical output unit:
> 0.0 : formatted data are written on WHAT(3) unit
              < 0.0 : unformatted data are written on |WHAT(3)| unit
                Values of |WHAT(1)|< 21 should be avoided (with the
                exception of +11).
                Default: WHAT(3) = 11.0 (standard output unit)"
Ciao
Andrea

> Dear users, it is me again,
> in my program I'm asking fluka a simple Energy (particle code 208)
> released output on a cilindrical body with the card USRBIN with an
> unformatted output unit.
> When it comes to results, i process them using usbsuw and then usbrea and
> i receive my data matrix but when it comes to the fluctuations, in the
> second part of the unit, i get always the same result 9.999 e-10 when the
> correspondin energy bin is different from zero otherwise i get zero as
> well.
> the card is :
> * Energy released by proton beam into OCO water cylinder (GeV/cm3)
> USRBIN 11.0 208.0 -19.0 +25.0
> +100.0cylinder
> USRBIN +0.0 +50.0 100.0 100.0 200.0&
>
> What do you thin the problem might be!?
> I mean, fluctuations should be different from on value to the other and
> not all identical....
> Thank you all!
> Antonello Carloni
>
>